Contact Us
- 0904.634.288, 024.8888.4288
- [email protected]
- Facebook Fanpage: http://fb.com/dauthau.info
- Messenger: http://m.me/dauthau.info
Contact Us
| No | List of product/services | Quantity | Unit | Description | Location | Specification | Note | Feature |
|---|---|---|---|---|---|---|---|---|
1 |
Breathing machine |
1 |
Machine |
General requirements; - Main machine manufacturing year: 2025 and later; - Machine quality: 100% new; - Meet ISO 13485 quality standards or equivalent; - Power supply: Suitable for voltage in Vietnam; - Working environment conditions:; Maximum temperature: ≥ 30°C; Maximum humidity: ≥ 70 %; Supply requirements; Anesthesia machine with accessories: 01 machine; Gas power cord set: 01 set; Flow sensor: 02 pieces; Isoflurane or Sevoflurane vaporizer: 01 bottle; CO2 absorption system with multiple-use soda lime tank: 01 set; Anesthetic gas release kit: 01 set; Accessories can be used from other manufacturers:; Multi-use adult mask: 01 piece; Anesthesia wire for adults, multiple use: 01 piece; Small size newborn or child mask for multiple use: 01 piece; Small sized neonatal or pediatric anesthesia wire for multiple use: 01 piece; Adult squeeze ball: 01 piece; Air compressor (purchased in Vietnam): 01 piece; User manual in English, Vietnamese, and quick user manual: 01 set; Technical requirements; General features:; - Anesthesia and ventilation support for patients from infants, children to adults; - The machine creates a ventilation volume (Vt) of 5 mL; - Machine used for low-flow anesthesia with fresh air flow from 0.1 liters/minute to 10 liters/minute; - The machine has an integrated auxiliary air exhaust system (ACGO); - Anesthesia machine with a minimum oxygen flow of 25 mL/minute; -The machine has interface options in many languages, including Vietnamese; - The monitoring screen is 7.5 inches in size |
Hoa Binh Regional Health Center |
|||
2 |
Monitor patient monitoring (5-7 parameters) |
2 |
Machine |
General requirements; Year of manufacture: Main equipment manufactured in 2025 or later, 100% new; Obtained certification: ISO 13485 or equivalent; Power supply: Suitable for voltage in Vietnam; Working environment conditions:; Maximum temperature: ≥ 30°C; Maximum humidity: ≥ 70 %; Supply requirements; Main machine: 01 unit; ECG cable: 01 piece; Disposable ECG electrode: 01 package; Blood pressure tube: 01 piece; Adult blood pressure cuff: 01 piece; Temperature measuring cable: 01 piece; Adult SpO2 measuring cable: 01 piece; IBP invasive blood pressure measurement set: 01 set; EtCO2 measuring set: 01 set; User manual in English, Vietnamese, and quick user manual: 01 set; Technical requirements; - Cooling system: fanless design; - Touch TFT LCD display screen: 12.1 inches with 1280 x 800 resolution; - Waveform display of up to 10 waveforms; - ECG:; + Scanning speed: 6.25, 12.5, 25, 50 mm/s, error ≤±10%; + Amplification: X0.125, X0.25, X0.5, X1, X2, X4, auto; - Heart rate:; + Heart rate range; Adults: 15~300 beats/minute; Children/Infants: 15~350 beats/minute; - Breathing rate; + Breathing rate range: 0 to 200 breaths/minute; - NIBP; + Method: Automatic oscillation; + Static measurement range: 0~300 mmHg (0kPa~40.0 kPa); - SPO2:; + Accuracy Children/Adults: ±2% (70~100%), Infants: ±3% (70~100%);; + SpO2 range: 0 to 100%; - Temperature:; + Technique: Thermistor; + Number of channels: 2 channels; + Temperature range: 0~50°C; - EtCO2; + Unit: mmHg, kPa; + Range (0 - 20.0kPa (at 760mmHg)): 0~150mmHg, 0vol%~19.7vol%; + Accuracy: ±1 bpm; - IBP:; + Number of channels: Maximum 2 channels; + Sensitivity: 5 μV/V/mmHg; + IBP range: -50 to 370mmHg; - Information transmission: connecting to CMS, HL7 |
Hoa Binh Regional Health Center |
|||
3 |
Endoscopic surgery system |
1 |
System |
General requirements must be met; manufactured from 2025 onwards, main machine origin from G7 countries; 100% new quality; meets ISO 13485 or equivalent quality standards; Working environment conditions: Maximum temperature: ≥ 30°C; Maximum humidity: ≥ 70%; Power supply: Suitable for voltage in Vietnam; Required components: Full HD image processor: 1 unit; Full HD endoscope camera head: 1 unit; Endoscopic cold light source: 1 unit; Light guide wire, 250 cm, 4.8 mm diameter: 1 unit; High-flow CO2 gas pump: 1 unit; Dedicated full HD endoscope monitor: 1 unit; Dedicated endoscope trolley with lockable drawer: 1 unit; High-frequency electrosurgical knife: 1 unit; User manual in English, Vietnamese, and quick start guide: 1 set; Technical Requirements; 1. Full HD Image Processor; Technical Features:; - Compatible with various types of camera heads and video endoscopes; - Has ≥ 2 different camera connection ports; - Image processing module supports:; + 180-degree image rotation; + Mirror image flip; + Picture-in-picture feature; - Adjustable image brightness; - Includes a filter for flexible fiber endoscopes; - Digital magnification ≥ 4 levels (maximum ≥ 2.0x); - Image recording/capture to USB; - Secure patient data system; - Resolution: ≥ 1920 x 1080 pixels; - Output signal ports: 2x DVI-D; - Has ≥ 4 USB 2.0 ports; - Minimum image capture in JPEG format; - Minimum video recording in MPEG-4 format; - Internal memory capacity: ≥ 50 GB; 2. Full HD camera head; - Image sensor: ≥ 1x1/3” CMOS; - Focal length of camera head f ≤ 24 mm; - Image resolution: ≥ 1920 x 1080 pixels; - ≥ 2 programmable function buttons on the camera head; - Can be immersed and autoclaved; - Compatible with endoscopes and instruments with standard connectors; - Weight: ≤ 130g; - Cable length: ≥ 3m; 3. Endoscopic cold light source; - Uses high-performance LED light source; - Lamp life: 30,000 hours; - Allows continuous brightness adjustment with a knob on the front of the device; - Has a warning mode when the device temperature increases and automatically stops operation when the temperature continues to rise too high; Technical specifications:; - Lamp type: High-performance LED; - Color temperature: 6400K; 4. Light guide wire; - Length: ≥ 250 cm; - Autoclaveable; 5. High-flow CO2 gas pump; - Touchscreen control or equivalent; - Maximum flow rate ≥ 40 liters/minute; - Maximum pressure ≥ 30 mmHg; - High flow mode available; - Displays minimum parameters: Pressure, flow rate, gas consumption; - Automatically adjusts flow rate according to instrument air resistance; - The device automatically remembers the settings from the last use and displays these parameters upon restart; 6. Dedicated Full HD endoscopy monitor; - LCD or IPS screen, LED backlight or equivalent; - Screen size: ≥ 27 inches; - Brightness: ≥ 300 cd/m2; - Contrast ratio: ≥ 1000:1; - Input signal Compatible with systems: DVI, SDI, or HDMI; - Output signal compatible with DVI; - Resolution: ≥ 1920 x 1080 full HD; 7. High-frequency electrosurgical knife; - Maximum power ≥ 350W; Memory capable of storing ≥ 10 modes; Automatic power consumption monitoring; Automatic fault detection and self-checking; ≥ 4 monopolar cutting modes and ≥ 3 monopolar hemostatic modes; ≥ 2 bipolar cutting modes and ≥ 3 bipolar hemostatic modes; 8. Endoscopy system trolley; - ≥ 3 equipment shelves; - Integrated ≥ 6 power outlets; - Endoscopy equipment trolley, moves on 4 double insulated wheels with wheel locks, has drawers and a CO2 cylinder holder. |
Hoa Binh Regional Health Center |
|||
4 |
Endoscopic Surgery Kit |
1 |
Set |
Meets general requirements; Production from 2025 onwards; 100% new quality; Meet ISO 13485 quality standards or equivalent; Technical characteristics; Technical list requirements (error no more than 5% compared to requirements or equivalent usage features):; Rigid prism scope, viewing direction 30 degrees, diameter 10 mm: 01 piece; Multi-function trocar valve, size 11 mm: 02 pieces; Trocar shell without valve, size 11 mm: 02 pieces; Trocar tip, smooth, sharp barrel, size 11 mm: 01 piece; Plain trocar, blunt trocar barrel, size 11 mm: 01 piece; Multi-purpose valve, size 6 mm: 02 pieces; Trocar shell without valve, size 6 mm: 02 pieces; Tip trocar, smooth bore, size 6 mm: 01 piece; Plain trocar, blunt trocar barrel, size 6 mm: 01 piece; Trocar tip rubber cap 11 mm: 01 piece; 6 mm trocar tip rubber cap: 01 piece; Lens reduction cap, size 11/5 mm: 01 piece; Clickline handle, plastic, no lock: 02 pieces; Clickline handle, plastic, with lock: 03 pieces; Clickline outer shell, size 5mm: 05 pieces; Clickline clamp forceps jaw with many small teeth: 01 piece; Clickline Kelly clamping and dissecting forceps jaw size 5 mm: 01 piece; Clickline clamp forceps jaw, with opening hole, does not cause damage: 01 piece; Clickline curved scissors blade, size 5 mm: 01 piece; Clickline hook scissors: 01 piece; Surgical and hemostatic electrode, L-shaped head: 01 piece; Suction irrigation tube, size 5 mm: 01 piece; TAKE-APART handle: 01 piece; Bipolar endoscopic instrument outer shell: 01 piece; Inner sheath of bipolar endoscope: 01 piece; Bipolar forceps: 01 piece; Unipolar high frequency cable. 300cm: 01 piece; Bipolar high frequency cable, 4 mm double pin: 01 piece; Plastic box for sterilization and storage of tools, perforated, transparent lid: 01 piece; Polymer hemostatic forceps, size L: 01 piece; Polymer hemostatic clip size L: 84 pieces |
Hoa Binh Regional Health Center |
|||
5 |
Electric birthing table |
2 |
Female |
General requirements; 100% new equipment, manufactured in 2025 or later; Meets ISO 13485 standards or equivalent; Power supply: Suitable for voltage in Vietnam; Working environment conditions:; Maximum temperature: ≥ 30°C; Maximum humidity: ≥ 70 %; Supply requirements; Bed body: 1 set; Mattress: 1 set; Electric push rod: 3 clusters; Armrests: 2 pieces; Leg support: 2 pieces; Fluid collection tray: 1 piece; Handheld controller: 1 piece; Power cord: 1 piece; Anesthesia curtain frame: 1 set; Handles: 2 pieces; Storage battery (backup battery): 1 set; User manual in English, Vietnamese, and quick user manual: 1 set; Technical requirements; Bed surface length: 1850 ± 20 mm; Bed width: 595 ± 20 mm; Side table size: 600 × 600 mm; Lowest height of bed surface: 710 ± 20 mm (electrically adjustable); Highest height of bed surface: 930 ± 20 mm (electrically adjustable); Bed tilt forward: ≥20° (electrically adjustable); Bed tilt backwards: ≥20° (electrically adjustable); Back plate lifting angle: 75° (electrically adjustable); Back plate lowering angle: 10°; Footrest: 90° horizontal rotation, removable |
Hoa Binh Regional Health Center |
|||
6 |
Newborn resuscitation table |
1 |
Female |
General requirements; Year of manufacture: Main equipment manufactured in 2025 or later, 100% new; Obtained certification: ISO 13485 or equivalent; Power supply: Suitable for voltage in Vietnam; Working environment conditions:; Maximum temperature: ≥ 30°C; Maximum humidity: ≥ 70 %; Supply requirements; - Main machine with 10.4 inch touch LCD screen; - Integrated heating lamp head: 01 piece; - Skin temperature sensor: 01 piece; - Integrated APGAR and CPR timer; - Examination lamp: 01 piece; - Rack and hanging pole: 01 piece; - Power cord: 01 piece; User manual in English, Vietnamese, and quick user manual: 01 set; Technical requirements; - Use far infrared heating bulbs; Better tissue penetration; Increase heat conversion efficiency; Reduce electricity consumption; Flexible rotating heating head; Horizontal rotation ±120°; Vertical tilt ±30°; Handles on both sides; - Integrated examination light:; Brightness adjustment 0–1500 lux; Fixed multiple angles; - 10.4 inch screen; Heating capacity 500W; Skin temperature setting range 34.0 – 38.0°C; Skin temperature accuracy ±0.3°C; Temperature display range: 20.0 – 42.0°C; Timer: Switchable APGAR/CPR mode; Examination light: LED 6W (2W × 3) |
Hoa Binh Regional Health Center |
|||
7 |
Obstetric monitor |
2 |
Machine |
General requirements; Year of manufacture: Main equipment manufactured in 2025 or later, 100% new; Obtained certification: ISO 13485 or equivalent; Power supply: Suitable for voltage in Vietnam; Working environment conditions:; Maximum temperature: ≥ 30°C; Maximum humidity: ≥ 70 %; Supply requirements; Main machine: 01 unit; Fetal heart rate monitoring probe: 01 piece; Probe for monitoring uterine contractions: 01 piece; Printing paper: 01 set; Battery: 01 piece; Rolling belt: 01 piece; Power cable: 01 piece; User manual in English, Vietnamese, and quick user manual: 01 set; Technical requirements; - Main machine protection level: IPX2; - Probe protection level IP68; - Display screen: 15.6 inches; - Resolution 1920 x 1080; - Fetal heart rate (FHR) monitoring probe; + Operating mode: pulsed Doppler ultrasound using auto-correlation algorithm (PW with auto-correlation); + Working frequency (1+/- 10%) MHz; + FHR measurement range: 50-240bpm; + Resolution: 1bpm; - TOCO uterine contraction monitoring probe; + TOCO range: 0-100%; + Nonlinear error: ≤ ±10%; + Resolution: 1%; - MFM & AFM (Manual and automatic fetal movement detection); - HIS connection data transmission: HL7 |
Hoa Binh Regional Health Center |
|||
8 |
Electrical syringe |
2 |
Female |
General requirements; - Manufactured in 2025 and later, 100% new; - Meets quality standards: ISO 13485 or CE or FDA certificate or equivalent certificate.; Power supply: Suitable for voltage in Vietnam; Supply requirements; - Main machine: 01 unit; - Accessory:; + Power cord: 01 piece; + Transmission pole clamp: 01 piece; User manual in English and Vietnamese, quick user manual: 01 set; Technical requirements; - Has a 4.3 inch color screen displaying parameters; - DC power source: Rechargeable battery inside the machine operates continuously for 12 hours when the mains power is lost; - There is an additional battery to support alarms in case of battery failure and power outage; - Can use syringe sizes: 5 mL, 10 mL, 20 mL, 30 mL, 50/60 mL (With 17 different syringe labels); - Select dose setting mode: Including modes mL/hour, µg/kg/minute, mg/kg/hour; - Injection speed setting range: 0.01 to 1200 mL/hour depending on syringe size |
Hoa Binh Regional Health Center |
|||
9 |
Transmission machine |
2 |
Female |
General requirements; - Manufactured in 2025 and later, 100% new; - Meets quality standards: ISO 13485 or CE or FDA certificate or equivalent certificate.; Power supply: Suitable for voltage in Vietnam; Supply requirements; - Main machine: 01 unit; - Accessory:; + Power cord: 01 piece; + Droplet counting sensor: 01 piece; + Transmission pole clamp: 01 piece; User manual in English and Vietnamese, quick user manual: 01 set; Technical requirements; - Has a 4.3 inch color screen displaying parameters; - DC power source: Rechargeable battery inside the machine operates continuously for 5 hours when the mains power is lost; - There is an additional battery to support alarms in case of battery failure and power outage.; - There are error lights and machine status; - Transmission speed:; 3 to 300mL/hour (for 20 drops/mL infusion lines); 1 to 100 mL/hour (for 60 drops/mL infusion lines); Installation step: 1 mL/hour |
Hoa Binh Regional Health Center |
|||
10 |
Portable surgical light (Led) |
2 |
Female |
General requirements; - Manufactured in 2025 and later, 100% new; - Meets quality standards: ISO 13485 or CE or FDA certificate or equivalent certificate.; Power supply: Suitable for voltage in Vietnam; Supply requirements; Lamp head: 1 set; Sterilization handle: 2 (1 standard and 1 spare); Stand (With braked wheels): 1 set; Switching power supply: 1 set; Tool set included with the lamp: 01 set; Instructions for use in English, Vietnamese and quick user manual: 01 set; Technical requirements; Number of LED bulbs: ≥ 48 bulbs; Brightness (Lux): 40,000 - 160,000; Color temperature (K): 3,000 – 6,700K; Bright area diameter (mm): 160-280; Brightness adjustment system: continuous; RA: ≥92; Temperature increase at the surgeon's head (℃): ≤1; Temperature increase in the surgical area (℃): ≤2; Color rendering index (CRI): ≥96; Color reproduction index: ≥97; Illumination depth (mm): ≥1200 |
Hoa Binh Regional Health Center |
|||
11 |
High frequency cervical cautery machine |
1 |
Machine |
General requirements; 100% new equipment, manufactured in 2025 or later; Equipment meets ISO 13485 standards, or CE or equivalent; Origin: OECD member; Power supply: Suitable for voltage in Vietnam; Supply requirements; Main machine: 01 unit; Double foot pedal: 01 piece; Silicone cooling pad: 01 piece; Disposable cooling pad: 01 piece; Knife handle with 2-button manual switch: 01 piece; Unipolar knife handle: 01 piece; Bipolar clamp: 01 piece; Blade set: 01 set; Instructions for use in English, Vietnamese and quick user manual: 01 set; Technical requirements; Characteristic:; Distinguish each working condition by sound and LED lights.; Stable output power using microprocessor.; When the device is turned off and when it is turned on again, the last set value is displayed by the flash memory.; Regime:; Fine cutting: 200W at 200 Om impedance; Mixed cutoff 1: 150W at 200 Om impedance; Mixed cut 2: 120W at 200 Om impedance; Contact coagulation: 80W at 200 Om impedance; Bipolar coagulation: 80W at 100 Om impedance; Free spray: 80W at 500 Om impedance; Main frequency: 400, 500 kHz; Protection mode: Class I, type CF |
Hoa Binh Regional Health Center |
|||
12 |
Newborn incubator |
1 |
Set |
General requirements; Year of manufacture: Main equipment manufactured in 2025 or later, 100% new; Obtained certification: ISO 13485 or equivalent; Power supply: Suitable for voltage in Vietnam; Working environment conditions:; Maximum temperature: ≥ 30°C; Maximum humidity: ≥ 70 %; Supply requirements; - Main machine: 01 unit; - Bracket and wheeled legs with fixed lock: 01 piece.; - Air temperature and humidity sensor: 01 piece; - Skin temperature sensor - reusable type: 02 pieces; - Mattress: 01 piece; - X-Ray tray: 01 pc.; - Weight scale: 01 set; - Height adjustment: 01 set; - Integrated SpO2 module: 01 set; - Jaundice phototherapy lamp: 01 set; - Examination light: 01 set; - Air filters: 04 pieces; User manual in English, Vietnamese, and quick user manual: 01 set; Technical requirements; + The incubator is made of 2-layer double walls, the front and back sides can be opened for easy operation.; + Has audio and visual alarm function; + Touch screen display; + The air temperature control range in the incubator is stable in the range of 20 ~ ≥ 370C. Can exceed thresholds up to 370C ~ ≥ 390C; + Accuracy: +/-10C; + Skin temperature control range via skin temperature sensor: 340C ~ ≥ 370C. Can control skin temperature beyond the threshold: 370C ~ ≥ 380C; + Accuracy: +/-30C; + Humidity control range in the incubator: ≤ 30~≥ 95% RH ±5%; + Adjustment system: Steam (boiling at 100 degrees Celsius); + Water tank volume 1,500 ml; + Noise level < 45 dBA; + Has an air filter; + Audio and visual alarms |
Hoa Binh Regional Health Center |
|||
13 |
Ventilator |
2 |
Machine |
General requirements; Year of manufacture: Main equipment manufactured in 2025 or later, 100% new; Certification: ISO 13485 or equivalent; Power supply: Suitable for voltage in Vietnam; Working environment conditions:; Maximum temperature: ≥ 30°C; Maximum humidity: ≥ 70%; Required supplies: Main unit: 1; Disposable bacterial filters: 2; Reusable adult breathing circuit board: 1; Reusable neonatal breathing circuit board: 1; Artificial lung: 1; Small size pediatric mask: 1; Adult breathing mask: 1; Neonatal flow sensor: 1; Oxygen sensor: 1; Support arm: 1; Oxygen supply hose: 1; Compressed air supply hose: 1; Power cord: 1; English and Vietnamese user manuals, quick start guide: 1 set; Technical requirements; - High-end resuscitation ventilator using a turbine system, convenient for intra-hospital transport; - Supports invasive ventilation with multiple modes; - Supports non-invasive ventilation with multiple modes; - 12.1-inch TFT capacitive touchscreen, 1280 × 800 pixel resolution; - Screen adjustment angle: horizontal 270°, vertical 45°; - Set oxygen concentration O2%: 21% ~ 100% vol.; - Tidal volume VT: adults 100 ~ 4000ml; children 20.0 ~ 300ml; neonates 2.0 ~ 100ml; - Breathing frequency f with dual tube: adults/children 1 ~ 100bpm; neonates 1 ~ 150bpm; With single tube: 1 ~ 120 bpm; - fSIMV: 1 ~ 60 bpm; - Tinsp: 0.10 ~ 10.00 seconds; - I:E ratio: 4:1 ~ 1:10; - Tslope: 0.00 ~ 2.00 seconds; - PEEP: double tube 0 ~ 50 cmH2O; single tube 0.0 ~ 15.0 cmH2O; - Pinsp: double tube 1 ~ 80 cmH2O; single tube 2.0 ~ 25.0 cmH2O; - Psupp: 0 ~ 80 cmH2O; Phigh: 0 ~ 80 cmH2O; Plow: 0 ~ 50 cmH2O; - Thigh: 0.10 ~ 30.00 seconds; 1. Tlow: 0.20 ~ 30.00 seconds; 2. Monitoring parameters; - O2% monitoring: 0% ~ 100% vol.; VT: 0.0 ~ 9999 ml.; - MV: adult/child 0.0 ~ 100.0 L/min; neonate 0.0 ~ 30.0 L/min.; - Frequency f: 0 ~ 200 bpm; I:E: 150:1 ~ 1:150; Tinsp: 0.00 ~ 60.00 seconds.; - PAW airway pressure: -60 ~ 120 cmH2O; PEEP: 0.0 ~ 120 cmH2O.; - Resistance: 0 ~ 600 cmH2O/(L/s); compliance: 0 ~ 300 ml/cmH2O.; 3. Safety warnings; - High VT warning: Adults 110 ~ 9999ml/OFF; Children 15 ~ 650ml/OFF; Neonates 3 ~ 600ml/OFF.; - Low VT warning: Adults 50 ~ 9995ml/OFF; Children 5 ~ 645ml/OFF; Neonates 1 ~ 595ml/OFF.; - High MV warning: Adults 0.20 ~ 100.0L/min; Children 0.20 ~ 60.0L/min; Neonates 0.02 ~ 30.0L/min; can be set OFF in NCPAP.; - Low MV warning: Adults 0.10 ~ 50.0L/min; Children 0.10 ~ 30.0L/min; Neonates 0.01 ~ 15.0L/min; can be set OFF in non-invasive ventilation. |
Hoa Binh Regional Health Center |
|||
14 |
C-PAP positive pressure ventilator |
3 |
Machine |
General requirements; - Year of manufacture: 2025 and later, 100% new; - Meets ISO 13485 and FSC quality standards (for main engines) or equivalent; Power supply: Suitable for voltage in Vietnam; Working environment conditions:; Maximum temperature: ≥ 30°C; Maximum humidity: ≥ 70 %; Supply requirements; Main machine with standard accessories for one machine, including: 01 set; Breathing tube set: 01 set; Temperature sensor and breathing air heating wire: 01 piece; Warming bottle: 01 set; Pressure vessel: 01 set; Machine support trolley: 01 piece; Comes with an air compressor; Need to fix and direct the breathing air wire: 01 piece; Set of air supply lines to the machine: 01 set; User manual in Vietnamese, English, and quick user manual: 01 set; Technical requirements; Setting parameters:; - FiO2: 21 – 100%; - Flow: 0 – ≥ 30 liters/minute; - PEEP: 0 – ≥ 10 cmH2O; - Breathing air temperature: ≤ 35 – ≥ 40 (̊C); Requirements for air supply; - Oxygen gas: From ≤ 280 kPa - ≥ 600kPa (or equivalent); - Compressed air: From ≤ 280 kPa - ≥ 600kPa (or equivalent); Display function; - LCD display screen; - Display: Set temperature, breathing air temperature; LED indicator; - Indicator that the machine is operating; - Alarm indication |
Hoa Binh Regional Health Center |
|||
15 |
Multi-parameter Monitor for Pediatrics - Newborns |
3 |
Machine |
General requirements; Year of manufacture: Main equipment manufactured in 2025 or later, 100% new; Obtained certification: ISO 13485 or equivalent; Power supply: Suitable for voltage in Vietnam; Working environment conditions:; Maximum temperature: ≥ 30°C; Maximum humidity: ≥ 70 %; Supply requirements; Main machine: 01 unit; ECG cable with 03 electrodes for newborns: 01 piece; Disposable ECG electrode: 01 package; Blood pressure tube: 01 piece; Newborn blood pressure cuff: 01 piece; Temperature measuring cable for children and newborns: 01 piece; Coiled SpO2 measuring cable for children and newborns: 01 piece; User manual in English, Vietnamese, and quick user manual: 01 set; Technical requirements; - Cooling system: fanless design; - Touch TFT LCD display screen: 12.1 inches with 1280 x 800 resolution; - Waveform display of up to 10 waveforms; - ECG:; + Scanning speed: 6.25, 12.5, 25, 50 mm/s, error ≤±10%; + Amplification: X0.125, X0.25, X0.5, X1, X2, X4, auto; - Heart rate:; + Heart rate range; Adults: 15~300 beats/minute; Children/Infants: 15~350 beats/minute; - Breathing rate; + Method: RA-LL impedance method; + Breathing rate range: 0 to 200 breaths/minute; + Apnea time: Adults: 10~60 seconds, resolution 5 seconds; Child/Infant: 10~40 seconds, resolution 5 seconds; - NIBP; + Method: Automatic oscillation; - SPO2:; + Resolution 1%; + Accuracy Children/Adults: ±2% (70~100%); Newborns: ±3% (70~100%);; - Temperature:; + Technique: Thermistor; + Number of channels: 2 channels; + Temperature range: 0~50°C; - Information transmission: connecting to CMS, HL7 |
Hoa Binh Regional Health Center |
|||
16 |
Electrical syringe |
2 |
Female |
General requirements; - Manufactured in 2025 and later, 100% new; - Meets quality standards: ISO 13485 or CE or FDA certificate or equivalent certificate.; Power supply: Suitable for voltage in Vietnam; Supply requirements; - Main machine: 01 unit; - Accessory:; + Power cord: 01 piece; + Transmission pole clamp: 01 piece; User manual in English and Vietnamese, and quick user manual: 01 set; Technical requirements; - Has a 4.3 inch color screen displaying parameters; - DC power source: Rechargeable battery inside the machine operates continuously for 12 hours when the mains power is lost; - There is an additional battery to support alarms in case of battery failure and power outage; - Can use syringe sizes: 5 mL, 10 mL, 20 mL, 30 mL, 50/60 mL (With 17 different syringe labels); - Select dose setting mode: Including modes mL/hour, µg/kg/minute, mg/kg/hour;; - Injection speed setting range: 0.01 to 1200 mL/hour depending on syringe size |
Hoa Binh Regional Health Center |
|||
17 |
High-end 4D and 5D ultrasound machines |
1 |
Machine |
General Requirements:
- Brand new, 100%, manufactured after 2025.
- Manufacturer: Meets ISO 13485 or equivalent quality standards.
- Product meets CE, FDA, or equivalent quality standards.
- Main unit origin: G7.
- Power supply: Compatible with Vietnamese voltage.
- Working environment conditions for the main unit:
- Maximum temperature: ≥ 30°C
- Maximum humidity: ≥ 70%
- Requirements:
- Main unit: 1 piece
- Display screen: 1 piece
- Screen rotation arm: 1 piece
- Touch control panel: 1 piece
- Linear frequency probe for vascular, small parts examination: 1 piece
- Convex frequency probe for general examination: 1 piece
- Vaginal frequency probe for gynecological examination: 1 piece
- 3D/4D fetal ultrasound probe with wide frequency range: 1 piece
- Clinical ultrasound software package including abdominal, vascular, small parts, musculoskeletal, obstetrics, gynecology, pediatrics, nephrology, cardiac function: 1 software package
- 4D obstetric ultrasound software: 1 software
- Full-screen ultrasound image extension software: 1 software
- Real-time image optimization software: 1 software
- DICOM network connection support: 1 package
- DVD drive: 1 piece
- Thermal black and white printer: 1 piece
- Color printer: 1 set
- Computer set: 1 set
- Online UPS ≥ 2kVA: 1 set
- Vietnamese, English, and quick start manual: 1 set
Technical Requirements:
1. Main unit:
- Synchronous design on trolley with 4 wheels and brake lock.
- ≥ 4 probe connection ports.
- Maximum system dynamic range: ≥ 350 dB.
- Maximum examination depth: ≥ 40 cm, depending on the probe and application.
- Gray scale: ≥ 256 levels.
- Total number of digital processing channels: ≥ 7,000,000 channels.
- Black and white image acquisition speed: ≥ 1900 images/second.
- Supports frequencies up to: ≥ 25 MHz.
- Hard drive capacity: ≥ 1 TB.
2. Display screen:
- LCD or equivalent, size ≥ 21.5 inches.
- Resolution: ≥ 1920 x 1080 pixels.
3. Control panel and user interface:
- Touch control panel, size ≥ 10.4 inches.
- ≥ 8 TGC adjustment sliders.
4. Probes:
- Electronic switching between probes.
- Users can customize image presets for each probe.
- Linear frequency probe for vascular, small parts examination:
- Frequency range: ≤ 5 – ≥ 12 MHz.
- Number of transducers: ≥ 192 transducers.
- Aperture: ≥ 50 mm.
- Application: vascular, small parts.
- Convex frequency probe for general examination:
- Frequency range: ≤ 2 – ≥ 5 MHz.
- Number of transducers: ≥ 128 transducers.
- Field of view (2D): ≥ 70 degrees.
- Application: abdominal, general.
- Supports biopsy guidance.
- Vaginal frequency probe for gynecological examination:
- Frequency range: ≤ 4 – ≥ 9 MHz.
- Number of transducers: ≥ 128 transducers.
- Field of view (2D): ≥ 160 degrees.
- Application: vaginal ultrasound.
- 3D/4D fetal ultrasound probe with wide frequency range:
- Frequency range: ≤ 2 – ≥ 8 MHz.
- Number of transducers: ≥ 192 transducers.
- Field of view (2D): ≥ 72 degrees.
- Application: obstetrics.
5. Ultrasound software and functions:
- Clinical ultrasound software package including abdominal, vascular, small parts, musculoskeletal, obstetrics, gynecology, pediatrics, nephrology, cardiac function.
- Measurement and calculation functions:
- General measurement and analysis function: Yes.
- Vascular measurement and analysis function: Yes.
- Small parts measurement and analysis function: Yes.
- Obstetrics measurement and analysis function: Yes.
- Gynecology measurement and analysis function: Yes.
- Pediatrics measurement and analysis function: Yes.
- Nephrology measurement and analysis function: Yes.
- Complete basic measurements: distance, circumference, volume, area, heart rate...
- Real-time Doppler analysis.
- Full-screen ultrasound image extension software: Yes.
- Real-time image optimization software: Yes.
- Continuously and in real time adjusts the gain and TGC to achieve balanced tissue brightness.
6. Image modes and display:
- 2D grayscale imaging.
- M-mode.
- M-mode Doppler color.
- Harmonic imaging.
- Real-time multi-directional imaging.
- Simultaneous M-mode 2D.
- Doppler color.
- Power and directional Doppler.
- Pulse Wave (PW) Doppler frequency: up to ≥ 19.8 kHz.
7. 2D (B-mode) features:
- Image inversion left/right, top/bottom.
- Gain control.
- Supports image acquisition speed ≥ 1900 frames/second.
- Up to ≥ 5 levels of noise reduction technology.
8. M-mode features:
- Can select scan speed.
9. Doppler features:
- Color Doppler: up to ≥ 9 selectable color bands.
- Color inversion.
- Color on animation and static.
- Color Doppler PRF frequency: up to ≥ 19.8 kHz.
- Power Doppler:
- High sensitivity mode for observing small vessels.
- Adjust flow density.
- Color on animation and static.
- Power Doppler PRF frequency: up to ≥ 19.8 kHz.
- Pulse Wave (PW) Doppler frequency: ≤ 0.2 – ≥ 34 kHz.
10. Cine image memory:
- 2D/Color data: ≥ 2000 images.
- Doppler and M mode data: up to ≥ 60 seconds.
11. Data storage:
- System storage capacity: ≥ 512 GB.
- Supports DVD drive.
12. Connectivity:
- USB port.
- Printer connection port.
- Wireless network, wired network, DICOM.
- DisplayPort or HDMI or equivalent. |
Hoa Binh Regional Health Center |
|||
18 |
Cervical Endoscope |
1 |
Machine |
General requirements; - 100% new machine, manufactured in 2025 or later.; - Manufacturer: Meets ISO 13485 quality standards or equivalent; Power supply: Suitable for voltage in Vietnam; Supply requirements; - Main camera: 01 piece; - Power cord: 01 piece; - Connection cable: 01 set; - Main camera bracket: 01 piece; - Handheld controller: 01 piece; - Foot pedal: 01 piece; - Medical endoscopy standard screen 22": 01 piece; - Screen support arm: 01 piece; - Computer + printer: 01 set; - Genuine cart: 01 piece; - HD video card: 01 piece; - Dedicated English software: 01 set; - English and Vietnamese user manual, and quick user manual: 01 set; - Dedicated Vietnamese software; Technical specifications; Main camera; - Sensor: 1/3” progressive scan CMOS; - Resolution: 2,100,000 Pixels; - Image size: 1920*1080; - Scanning system: Continuous scanning; - Frame rate: (1920*1080)/60P; - SNR: ≥50dB; - Output signal: Y/Cb/Cr4:2:2 via SDI video signal; - Horizontal resolution: ≥800 TV lines; Specialized trolley:; - Height: 84.6cm; - Shelf size: 59.6 * 48.8cm; - Control panel size: 59.6 * 35.5cm; - Weight: 45kg; - Material: aluminum alloy. Has anti-corrosion coating; - Wheels: multi-purpose without making noise, size 4"; - Flexible screen mounting arm can hang screens of sizes: 32", 24" and 22"; Specialized screen:; - High performance I/P conversion; - Bright separation, deep compensation; - Environmentally friendly LED backlight; - Dynamic dark area color reproduction technology; - Over Drive mode speeds up response and eliminates screen blur; - Instant digital image enhancement technology; - Unique automatic white balance adaptive technology; - Gamma value is adjusted one by one before leaving the factory; - Personalized three basic color adjustment rights; - Customize screen ratio adjustment between different modes |
Hoa Binh Regional Health Center |
|||
19 |
Immunosuppressor |
1 |
Machine |
General requirements; 100% new machine, manufactured in 2025 or later.; The factory meets quality standards: ISO 13485 or equivalent; Power supply: Suitable for voltage in Vietnam; Working environmental conditions: Maximum temperature: ≥ 30°C Maximum humidity: ≥ 70 %; Supply requirements; Main machine and standard accessory set: 01 set; Computer set: 01 set; Handheld barcode scanner: 01 set; UPS: 01 unit; Trial chemical set: 01 set; Set of instructions for use in English, Vietnamese, and quick user manual: 01 set; Technical requirements; Type: immune analyzer; Capacity: ≥ 60 tests/hour; Principle of discovery: ....; Machine functions: random access function, continuous sample loading, emergency testing; Type of sample tube: microcup and blood sample tube; Sample volume used for testing: 5 – 150 µl; Specimen suction needle: has the function of detecting blood clots and detecting liquid levels; Has the function of identifying patient samples by barcode: Codabar, code 39, code 128, ITF; Sample type: human serum, plasma, urine; Reaction tube loading capacity: 500 reaction tubes, with continuous reaction tube loading function; Reagent conveyor temperature ranges from: 2 – 8 °C; Volume of R1/R2 reagent consumed: 20 – 150 µl, depending on the test; Reagent aspiration needle: Has the function of detecting reagent level; Has the function of identifying reagents by barcode; Has the function of predicting test result return time; Has automatic dilution and automatic rerun mode; Calibrator is included in the reagent kit; Capable of continuously loading samples and reagents |
Hoa Binh Regional Health Center |
|||
20 |
Electrolyte testing machine |
1 |
Machine |
General requirements; 100% new machine, production year from 2025 onwards; - Meets quality standards: ISO 13485 or equivalent; Power supply: Suitable for voltage in Vietnam; - Working environment conditions: Maximum temperature: ≥ 30°C Maximum humidity: ≥ 70 %; Supply requirements; - Main machine Model and standard accessory set: 01 set; - Thermal printer integrated on the machine: 01 set; - Thermal printing paper: 02 rolls; - Handheld barcode scanner: 01 piece; - Power cord: 01 set; - Trial chemical set includes: 01 set; - User manual in English and Vietnamese, and quick user manual: 01 set; Technical requirements; - Electrolyte analyzer - Test parameters: 5 parameters (K+, Na+, Cl-, Ca2+, pH); - Speed:; - Using multi-biosensor technology - No need for electrode maintenance; - Testing principle: Ion selective electrode; - Has an air bubble sensor to help detect and warn of air bubbles in the fluid line; - Sample type: Serum, urine, whole blood; - Built-in thermal printer on the machine; - RS232 and USB ports connect to computers and barcode readers |
Hoa Binh Regional Health Center |
|||
21 |
Blood coagulation testing machine |
1 |
Machine |
General requirements; - 100% new machine, manufactured in 2024 or later.; - Meets quality standards: ISO 13485 or equivalent; Power supply: Suitable for voltage in Vietnam; - Working environment conditions: Maximum temperature: ≥ 30°C Maximum humidity: ≥ 70 %; Supply requirements; - Main machine and standard accessory set: 01 set; - Computer set (purchased in Vietnam): 01 set; - UPS (purchased in Vietnam): 01 set; - External barcode scanner (purchased in Vietnam): 01 set; - Instructions for use in English and Vietnamese: 01 set; - Trial chemical set: 01 set; - User manual in English and Vietnamese, and quick user manual: 01 set; Technical requirements; - Fully automatic analyzer, with emergency sample run function; - Analysis speed; - Testing method + Based on the principles of mechanical viscosity, colorimetry, immunology + Not affected by hemolytic samples, jaundice, etc. + Multi-wavelength detection technology allows simultaneous measurement on multiple channels with different methods; - Measured parameters: TT, APTT, PT, FIB, protein C, Protein S, D-Dimer, FDP, AT-III, endogenous/exogenous coagulation factors; - There are 2 suction needles, with separate sample and reagent needles; - Sampling volume: 5 – 100 µl (depending on test); - Sample location: supports running STAT samples; - Automatically re-dilute and re-run the test with abnormal samples.; - Pre-calibration function helps reduce the cost of standard plasma and reagents.; - Able to connect to an external barcode reader; - Connection: + RS-232, RJ45, USB port + Supports 2-way LIS/HIS connection + Can connect to external printers, external barcode scanners; 21 categories |
Hoa Binh Regional Health Center |
Select columns to display
If you are still not our member, please sign up.You can use your account on DauThau.info to log in to Dauthau.Net, and vice versa!
Only 5 minutes and absolutely free! Your account can be use on all over our ecosystem, including DauThau.info and DauThau.Net.
Detailed instructions on how to deposit money and redeem points on DauThau.info
When does contract negotiation fail?
[VIDEO] Guide for bid solicitors to Extend the Bid Submission Deadline
[VIDEO] Guide to Evaluating Bid Documents on the New National Bidding Network System
[VIDEO] Guide to Posting Tender Notices for PPP Projects
[VIDEO] Guide to Posting Prequalification Notices for PPP Projects
[VIDEO] Guide to Responding to Clarification Requests for e-HSMST (E-Prequalification Invitation Documents) of PPP Projects on the New National Bidding Network System
[VIDEO] Guide to Uploading the Shortlist for PPP Projects
Similar goods purchased by Investor Trung tâm Y tế khu vực Hòa Bình:
No similar goods purchased by the Investor were found.
Similar goods purchased by other Investor excluding Trung tâm Y tế khu vực Hòa Bình:
No similar goods have been found that other Investor have purchased.
Similar goods for which the bidding results of the Ministry of Health have been announced:
No similar goods have been found in the bidding results of the Ministry of Health.