| No | List of product/services | Quantity | Unit | Description | Location | Specification | Note | Feature |
|---|---|---|---|---|---|---|---|---|
1 |
Vaccine storage cabinet |
1 |
Female |
Vaccine storage cabinet
Model: RCW50EG
Brand: Dometic- Electrolux, Luxembourg; Phone number: 3040483
Maintenance and replacement of components for
- Check the tightness of the gaskets
- Monitor and check the condenser system
- Check the cooling level
- Clean the entire cooling system and electronic circuit board.
- Check the machine's operating functions.
- Measurement calibration
Automatic temperature monitoring device for cold chain with external sensor wire
Technology compatibility with the Investor's existing machines:
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
After repairing, replacing components and maintaining, the Equipment is inspected and measured by a licensed unit of a competent licensing agency. |
Duc Xuan TYT |
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation. Meets specifications. Meet the user needs of beneficiary units |
||
2 |
Vaccine storage cabinet |
1 |
Female |
Vaccine storage cabinet
Model: TCW 3000
Brand: Dometic- Electrolux, Luxembourg; Phone number: 8011140
Maintenance and replacement of components for
- Check the tightness of the gaskets
- Monitor and check the condenser system
- Check the cooling level
- Clean the entire cooling system and electronic circuit board.
- Check the machine's operating functions.
- Measurement calibration
Automatic temperature monitoring device for cold chain with external sensor wire
Technology compatibility with the Investor's existing machines:
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
After repairing, replacing components and maintaining, the Equipment is inspected and measured by a licensed unit of a competent licensing agency. |
Huyen Tung TYT |
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation. Meets specifications. Meet the user needs of beneficiary units |
||
3 |
Vaccine storage cabinet |
1 |
Female |
Vaccine storage cabinet
Model: RCW50EG
Brand: Dometic- Electrolux, Luxembourg; Phone number: 3040481
Maintenance and replacement of components for
- Check the tightness of the gaskets
- Monitor and check the condenser system
- Check the cooling level
- Clean the entire cooling system and electronic circuit board.
- Check the machine's operating functions.
- Measurement calibration
Automatic temperature monitoring device for cold chain with external sensor wire
Technology compatibility with the Investor's existing machines:
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
After repairing, replacing components and maintaining, the Equipment is inspected and measured by a licensed unit of a competent licensing agency. |
CHC Nguyen Thi Minh Khai |
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation. Meets specifications. Meet the user needs of beneficiary units |
||
4 |
Vaccine storage cabinet |
1 |
Female |
Vaccine storage cabinet
Model: RCW50EG
Brand: Dometic- Electrolux, Luxembourg; Phone number: 3040473
Maintenance and replacement of components for
- Check the tightness of the gaskets
- Monitor and check the condenser system
- Check the cooling level
- Clean the entire cooling system and electronic circuit board.
- Check the machine's operating functions.
- Measurement calibration
Automatic temperature monitoring device for cold chain with external sensor wire
Technology compatibility with the Investor's existing machines:
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
After repairing, replacing components and maintaining, the Equipment is inspected and measured by a licensed unit of a competent licensing agency. |
Song Cau TYT |
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation. Meets specifications. Meet the user needs of beneficiary units |
||
5 |
Vaccine storage cabinet |
1 |
Female |
Vaccine storage cabinet
Model: RCW50EG
Brand: Dometic- Electrolux, Luxembourg; Machine number: 2500184
Maintenance and replacement of components for
- Check the tightness of the gaskets
- Monitor and check the condenser system
- Check the cooling level
- Clean the entire cooling system and electronic circuit board.
- Check the machine's operating functions.
- Measurement calibration
Automatic temperature monitoring device for cold chain with external sensor wire
Technology compatibility with the Investor's existing machines:
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
After repairing, replacing components and maintaining, the Equipment is inspected and measured by a licensed unit of a competent licensing agency. |
Nong Thuong TYT |
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation. Meets specifications. Meet the user needs of beneficiary units |
||
6 |
Vaccine storage cabinet |
1 |
Female |
Vaccine storage cabinet
Model: RCW50EG
Brand: Dometic- Electrolux, Luxembourg; Phone number: 3040474
Maintenance and replacement of components for
- Check the tightness of the gaskets
- Monitor and check the condenser system
- Check the cooling level
- Clean the entire cooling system and electronic circuit board.
- Check the machine's operating functions.
- Measurement calibration
Automatic temperature monitoring device for cold chain with external sensor wire
Technology compatibility with the Investor's existing machines:
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
After repairing, replacing components and maintaining, the Equipment is inspected and measured by a licensed unit of a competent licensing agency. |
Xuat Hoa TYT |
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation. Meets specifications. Meet the user needs of beneficiary units |
||
7 |
Vaccine storage cabinet |
1 |
Female |
Vaccine storage cabinet
Model: RCW50EG
Brand: Dometic- Electrolux, Luxembourg; Phone number: 3040490
Maintenance and replacement of components for
- Check the tightness of the gaskets
- Monitor and check the condenser system
- Check the cooling level
- Clean the entire cooling system and electronic circuit board.
- Check the machine's operating functions.
- Measurement calibration
Automatic temperature monitoring device for cold chain with external sensor wire
Technology compatibility with the Investor's existing machines:
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
After repairing, replacing components and maintaining, the Equipment is inspected and measured by a licensed unit of a competent licensing agency. |
Phung Chi Kien CHC |
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation. Meets specifications. Meet the user needs of beneficiary units |
||
8 |
Vaccine storage cabinet |
1 |
Female |
Vaccine storage cabinet
Model: RCW50EG
Brand: Dometic- Electrolux, Luxembourg; Phone number: 3040482
Maintenance and replacement of components for
- Check the tightness of the gaskets
- Monitor and check the condenser system
- Check the cooling level
- Clean the entire cooling system and electronic circuit board.
- Check the machine's operating functions.
- Measurement calibration
Automatic temperature monitoring device for cold chain with external sensor wire
Technology compatibility with the Investor's existing machines:
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
After repairing, replacing components and maintaining, the Equipment is inspected and measured by a licensed unit of a competent licensing agency. |
Duong Quang TYT |
Equipment components after being replaced and installed work well. Fix errors. Maintain machine operation. Meets specifications. Meet the user needs of beneficiary units |
||
9 |
Ceiling-mounted operating table |
1 |
Female |
Ceiling-mounted operating table
Model: ROL-501C
Machine number: ROL-501C10017
Adjust the slope, tighten the screws at the joints
Check the functions and parameters of the device
- Clean the machine
- Clean the main power board and power supply board
- Check and clean connectors and connection cables
Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
Suitable for the investor's existing technology. |
Bac Kan Medical Center |
|||
10 |
Electric scalpel |
2 |
Female |
Electric scalpel
Model: XCELL350MCDSC
Phone number: 7171-05/13
Model 16-2000-300HF
Machine number: 30910
- Reusable electric scalpel mass plate, vulcanized rubber material and accessories for electric scalpels
Check the functions and parameters of the device
- Clean the machine
- Clean the main power board and power supply board
- Check and clean connectors and connection cables
- Calibrate measurement parameters
Replace components for machines that match the technology with the Investor's existing machines:
After replacing, repairing and installing the machine, the machine works well. Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
11 |
Sterilized steamer |
1 |
Female |
Autoclave for sterilization
Model:TC409
Model: 1700981
-Temperature controller.
- Heating rod
- Start from
Check the functions and parameters of the device
- Clean the machine
- Clean the main power board and power supply board
- Check and clean connectors and connection cables
- Calibrate measurement parameters
Replace components for machines that match the technology with the Investor's existing machines:
After replacing, repairing and installing the machine, the machine works well. Fix errors. Maintain machine operation to meet technical specifications. Meets the needs of the user department.
Suitable for the investor's existing technology. |
Bac Kan Medical Center |
|||
12 |
Cast table |
1 |
Female |
Cutting stainless steel to process pipes with tightening straps.
Tailored to each patient |
Bac Kan Medical Center |
|||
13 |
Hand washing system |
1 |
System |
Hand washing system
Model: CHS-Matic Korea;
Machine number: Matic 0100085;
Year of manufacture: 2013
Automatic soap dispenser: 01 pc
Maintenance and replacement of components for Korean CHS-Matic hand washing system
Check the functions and parameters of the device
- Cleaning of equipment
- Clean the electrical circuit board
- Check the cleanliness of the control panel
- Check the cleanliness of function keys.
- Coarse filter: 02 pieces
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
14 |
Dental examination chair system |
1 |
Female |
Dental examination chair system
Model: Acutts 9000XL
Year of manufacture: 2013
Machine number: AC1301030
Fast dental handpiece with LED
(Handpiece with 1 water jet without light): 01 piece
Manufacturer: Nakanashi INC
Country of Manufacture: Japan
Maintenance and replacement of components for dental chairs Acutts 9000XL dental chair
- Maintain, clean, and descale the water supply and compressed air systems for the machine
- Maintenance and descaling of compressed air machines. Test pressure parameters, automatic switching and closing relays
- Check, clean, calibrate control relays, air valves, and water valves. Clean and anti-mold coating electrical circuits and connectors
- Recalibrate voltage levels for lights, relays, valves...
- Maintenance, specialized lubrication and speed calibration for fast and slow handpieces.
- Maintain and add lubricating oil and hydraulic oil to the patient bed lifting system
- Maintain and clean water pipes and calibrate the accompanying elevator
- Maintain and clean the included polymer light.
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
15 |
COMEN patient monitor |
1 |
Female |
COMEN patient monitor
Model: Star 8000E
Year of manufacture: 2023
- Blood pressure cuff: 01 piece
Maintenance and replacement of components for COMEN Star 8000E patient monitor
Check the functions and parameters of the device
- Clean the machine
- Clean the main power board and power supply board
- Check and clean connectors and connection cables
- Calibrate measurement parameters
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
16 |
High-frequency routine X-ray system |
1 |
System |
High-frequency routine X-ray system
Model: SHIMAZU
Year of manufacture: 2010
Maintenance for SHIMAZU high-frequency routine X-ray system
Shimadzu X-ray machine system, including:
1. Check the UD150L-30V high-frequency beam source control cabinet:
- Check the input/output voltage of the control arrays
- Check the installation parameters
- Check and handle the jacks and cables connecting to the control cabinet
- Check the mA control circuit
- Check kV control circuit
- Check and handle control buttons (if there is a problem)
- Check the rotating anode
- Check the beam bulb voltage
- Check the shadow line when shooting
- Check the operation of the frequency inverter circuit (Inverter)
- Check the error reporting system
- Check the values displayed on the control cabinet surface
2. Maintenance of high voltage cables:
- Check the heat generation of high pressure
- Check the coolant oil for the ray ball
3. Check the shooting table and ball column:
- Check the source arrays of tables and columns.
- Check and handle the sliding and rotating movements of the beam to be smooth.
- Check and handle the movement of the entire ball column sliding on the floor smoothly.
- Re-align DIAFRAM to be concentric.
- Check the operation of BUCKY shooting table. Check the central beam field.
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
17 |
Color ultrasound machine |
1 |
Female |
Color ultrasound machine
Acuson Germany
Model: Acuson NX2
Year of manufacture: 2018
Machine number:
(21)522619
Maintenance for German Acuson color ultrasound machine
Check the functions and parameters of the device
- Clean the machine
- Clean the main power board, power supply board, and control boards
- Check and clean connectors and connection cables
- Check the cleanliness of the probe
- Calibrate measurement parameters
Measurement verification
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
18 |
Cardisuny 3-channel ECG machine |
1 |
Female |
3-channel ECG machine
Model: Cardisuny C120
Japan
Phone number: 06104008
Electrocardiography cable for medical use
ECG cables connect from the 6-rod ECG machine with 6 chest electrodes and 4 limb electrodes
Maintenance and replacement of components for Japanese Cardisuny C120 3-channel ECG Machine
- Clean the entire device including electrodes, cables, and thermal print heads
- Operate, test and calibrate, install noise filters (if necessary)
- Fix minor errors (if any)
Measurement verification
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
19 |
Biomicroscopy for eye examination |
1 |
Female |
Biomicroscopy for eye examination
Model: SM-10N
Phone number: 0812239
- Shadows
- Belt raises and lowers the table
Maintaining furniture, cleaning mold, and drying
- Fix minor errors (if any)
Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
20 |
Biological safety cabinets |
1 |
Female |
Biological safety cabinet
Model: BSC-1100IIA2-X
Year of manufacture: 2020
- UV ultraviolet lamp
Maintenance and replacement of components for BSC-1100IIA2-X Safety Cabinet
- Clean all equipment with specialized disinfectant chemicals
- Replace microbiological filters (filters provided by customer)
- Check operations and fix minor errors (if any)
- Measurement verification
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
21 |
Pipette 4u |
3 |
Female |
Pipette 4u
Model: 4U
Year of manufacture: 2010
Pipette 4u Measurement Maintenance and Calibration
-Clean all equipment with specialized disinfectant chemicals
- Check operations and fix minor errors
- Measurement calibration
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
22 |
Steamer |
1 |
Female |
Steamer
Model: BKT-18L
Year of manufacture: 2017
Maintenance and Calibration of Autoclave, BKT-18L
Check the functions and parameters of the device
- Check and clean anti-burning
- Clean the machine
- Measurement calibration
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
23 |
Heat moisture |
3 |
Female |
Thermometer and hygrometer
Model: HE-173
Year of manufacture: 2015
Machine number: HE17XF1305; HE17XF1343;
HE17XF1407;
Maintenance and replacement of components for HE-173 Thermometer and Hygrometer
-Replace the Battery
- Inspection
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
24 |
Vaccine storage cabinet |
1 |
Female |
Vaccine Storage Cabinet
Model: TCW 4000AC
Year of manufacture: 2019
Phone number: 1082672
- Replace condenser and condenser components.
- Compressor for R600 Gas Cabinet
- Temperature controller
Maintenance and replacement of components for
- Check the tightness of the gaskets
- Monitor and check the condenser system
- Check the cooling level
- Clean the entire cooling system and electronic circuit board.
- Check the machine's operating functions.
- Measurement calibration
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
25 |
Fabric dryer |
1 |
Female |
Fabric dryer
Model: STI 34S-E USA
Replace the temperature controller and timer of the electromechanical system. Reverse timer and relay
Maintenance and replacement of components for American STI 34S-E fabric dryer
Check the functions and parameters of the device
- Cleaning the washing drum
- Check the discharge pump system
- Clean and check the keyboard
- Clean and check the control panel
- Lubricate the shaft and moving joints
- Replace 50A atomat: 01 piece
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
26 |
Industrial washing machine |
1 |
Female |
Industrial Washing Machine
Model: RC30 Italy
Washing machine belt
Maintenance and replacement of components for Italian RC30 Industrial Washing Machine
Check the functions and parameters of the device
- Cleaning the washing drum
- Check the discharge pump system
- Clean and check the keyboard
- Clean and check the control panel
- Lubricate the shaft and moving joints.
- Val water
Replacement parts, repair, and installation of the machine in good working order. Maintain machine operation to meet technical specifications. Suitable for the investor's existing technology. |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
||
27 |
Wastewater treatment system |
1 |
Female |
Replace the 3-phase 100A circuit breaker
Submersible wastewater pump: 01 piece
Wiring |
Bac Kan Medical Center |
Replacement components suitable for existing technology are available to the investor's beneficiary unit. The equipment works well. ensure the professional operation process of the beneficiary unit |
Downloading files directly on the new Public Procurement System requires a computer using Windows operating system and installing Client Agent software (Linux and MacOS operating systems cannot install Client software yet). Therefore, to be able to download files on smartphones, tablets or computers using operating systems other than Windows, you need to use DauThau.info software to download files.
To download, please Login or Register
Select columns to display
If you are still not our member, please sign up.You can use your account on DauThau.info to log in to Dauthau.Net, and vice versa!
Only 5 minutes and absolutely free! Your account can be use on all over our ecosystem, including DauThau.info and DauThau.Net.
Detailed instructions on how to deposit money and redeem points on DauThau.info
When does contract negotiation fail?
[VIDEO] Guide for bid solicitors to Extend the Bid Submission Deadline
[VIDEO] Guide to Evaluating Bid Documents on the New National Bidding Network System
[VIDEO] Guide to Posting Tender Notices for PPP Projects
[VIDEO] Guide to Posting Prequalification Notices for PPP Projects
[VIDEO] Guide to Responding to Clarification Requests for e-HSMST (E-Prequalification Invitation Documents) of PPP Projects on the New National Bidding Network System
[VIDEO] Guide to Uploading the Shortlist for PPP Projects
Similar goods purchased by Investor Trung tâm Y tế Bắc Kạn:
No similar goods purchased by the Investor were found.
Similar goods purchased by other Investor excluding Trung tâm Y tế Bắc Kạn:
No similar goods have been found that other Investor have purchased.
Similar goods for which the bidding results of the Ministry of Health have been announced:
No similar goods have been found in the bidding results of the Ministry of Health.