Purchasing a Radiofrequency Ablation (RFA) Tumor Treatment System

Find: 08:11 27/08/2026

Bid info

Project name/purchase estimate
Purchasing a Radiofrequency Ablation (RFA) Tumor Treatment System
Bid name
Purchasing a Radiofrequency Ablation (RFA) Tumor Treatment System

Basic information of the request for quotation

Request for quotation code
RQ2600035606-00
Status
Posted
Name of request for quotation
Purchasing a Radiofrequency Ablation (RFA) Tumor Treatment System
Address of the unit requesting a quote
Address of the unit requesting a quote (After merge)
Doi Can Street People's Group, Xã Vĩnh Tường, Province Phú Thọ
Categorize quotation requests
Bidding package for purchasing chemicals, testing supplies, and medical equipment

Contact information of the person responsible for receiving the quotation

Full Name
Ngô Thị Hồng Thắng
Position
Head of Department of VTTBYT
Phone number
Address
Department of Medical Equipment - Vinh Tuong Regional Medical Center - Yen Cat Village, Vinh Tuong Commune, Phu Tho Province

How to receive quotation

Form of receiving quotation
Offline
Location of request for quotation
http://muasamcong.mpi.gov.vn
Quotation acceptance time
Since 27/08/2026 08:10 To date 07/09/2026 17:00
Validity period of quotation
90 Day kể từ ngày 07/09/2026 17:00
Request for quotation
 

Content of request for quotation

Filter data

data/page
No List of product/services Quantity Unit Description Location Specification Note Feature
1
HT performed tumor ablation using radiofrequency ablation (RFA).
1
HT
A. General Requirements: 100% New Product, Manufacturing Year from 2026 onwards, Power Supply: 220V/50Hz or suitable for use in Vietnam. B. Expected Configuration Requirements: 1. General Requirements: * Utilizes Radiofrequency (RF) ablation technology for treating lesions/tumors as indicated by medical professionals. * Capable of being used in various procedures, including percutaneous interventions, endoscopy, or surgery, as designed and specified by the manufacturer. * Includes monitoring functions for parameters during ablation, such as power, current, impedance, temperature, and duration, or equivalent parameters, to ensure treatment control. * Provides alerts or detects neutral electrode connection status and abnormal conditions during operation. * Capable of storing, viewing, or exporting information, logs, or equivalent data from the procedure. * Meets applicable medical device safety and electromagnetic compatibility standards or equivalent standards. * Manufacturer has a suitable quality management system for medical devices, such as ISO 13485 or equivalent. * Device has appropriate legal documentation according to current regulations for medical equipment in Vietnam. 2. Main Machine: * Includes an integrated RF generator and control system. * Features a display screen for easy parameter setting and monitoring during treatment. * Capable of using 1 to 3 electrodes, as designed by the system. * Offers both automatic or manual ablation modes. * Allows control or adjustment of one or more parameters during ablation, such as power, temperature, duration, impedance, or equivalent parameters. * Includes a hemostasis function when withdrawing the electrode, if designed by the manufacturer. * RF Power: Suitable for treatment purposes and electrode design of the system. * Frequency: Suitable for device design and intended use. * Includes a function to monitor impedance during ablation. * Allows setting of RF transmission time and automatic stopping when the set time is complete or when termination conditions are met. * Provides alerts if abnormal conditions are detected, such as impedance, temperature, electrode connection, or neutral electrode status. 3. Cooling Pump: * Includes a cooling pump suitable for the ablation electrode with a cooling system. * Capable of providing stable cooling during procedures. * Flow rate appropriate for the requirements of the electrode and system. 4. RF Ablation Electrodes: * The system is capable of using electrodes suitable for treatment needs, including at least: * Fixed heat source ablation electrode; * Electrodes with different heat source head sizes; * Multi-electrode array electrodes, if supported by the system; * Bipolar electrodes, if supported by the system; * Electrodes with internal cooling systems, compatible with the cooling pump. * Electrodes must have appropriate size, working length, and heat source head length for the intended treatment of lesions or sizes. 5. Neutral Electrodes: * Includes neutral electrodes suitable for the system and electrode. * Provides a function to recognize/monitor neutral electrode connection status or equivalent safety functions. * Initial supply quantity: Suitable for device configuration. 6. Foot Pedal: * Includes a foot pedal to control the device. * Minimum Start/Stop function or equivalent. 7. Accessories: * Provides all standard accessories to enable normal operation of the device, including but not limited to: Cooling pump; Electrodes according to the manufacturer's standard configuration; Neutral electrodes; Coolant tubing; Foot pedal; Other standard accessories as designed by the manufacturer; Dedicated cart or equivalent, suitable for placing the machine and accessories; User manual in Vietnamese and English. The supplier may offer products with different configurations and specifications than those listed above if they can demonstrate that the products meet equivalent or better functionality, performance, safety, and application range.
TTYT KV Vĩnh Tường (likely an abbreviation for a specific entity)
Approval decision information
Approval date
26/08/2026
Approval decision code
894/TTYT-VTTBYT
Approval decision file

Downloading files directly on the new Public Procurement System requires a computer using Windows operating system and installing Client Agent software (Linux and MacOS operating systems cannot install Client software yet). Therefore, to be able to download files on smartphones, tablets or computers using operating systems other than Windows, you need to use DauThau.info software to download files.
To download, please Login or Register

Files with this icon are not supported by the public procurement system for direct download in normal mode. Visit DauThau.info and select Quick Download mode to download the file if you do not use Windows with Agent. DauThau.info supports downloading on most modern browsers and devices!
In case it is discovered that the BMT does not fully attach the E-HSMT file and design documents, the contractor is requested to immediately notify the Investor or the Authorized Person at the address specified in Chapter II-Data Table in E- Application form and report to the Procurement News hotline: 02437.686.611
Utilities for you
Follow quote requests
The Track Request for Quotation function helps you quickly and promptly receive email notifications of changes to Request for Quotation "Purchasing a Radiofrequency Ablation (RFA) Tumor Treatment System". Our software will track the requests for quotations that you have registered for you. When there is any change, it will immediately send you an email notification.
Receive similar quote requests via email
To be one of the first to receive similar Requests for Quotations by email: "Purchasing a Radiofrequency Ablation (RFA) Tumor Treatment System" as soon as they are published, subscribe to DauThau.info's VIP 8 package.
Views: 0

Select columns to display


You did not use the site, Click here to remain logged. Timeout: 60 second