| No | List of product/services | Quantity | Unit | Description | Location | Specification | Note | Feature |
|---|---|---|---|---|---|---|---|---|
1 |
Ultrasound |
1 |
Machine |
I. General Requirements:
- Must be manufactured from 2026 onwards, 100% new
- Main machine origin: OECD countries
- Meet quality standards: ISO 13485, FDA or equivalent
- Power supply compatible with Vietnamese power grid: 100 – 240 VAC ~ 50/60 Hz
- Working environment:
+ Operating temperature: ≥ 30°C
+ Operating humidity: ≥ 70%
II. Basic Configuration:
- Main machine: 1 unit
- 3 probes, including:
- Convex multi-frequency probe: 1 unit
- Linear multi-frequency probe: 1 unit
- Multi-frequency cardiac probe: 1 unit
- Online UPS ≥ 1KVA: 1 unit
- Computer system, including:
+ Desktop computer, processor: ≥ Core i5, RAM ≥ 8GB, SSD: ≥ 512 GB
+ HD monitor: ≥ 21.5 inch
- Color printer with water-based ink: ≥ 4 colors
- Monochrome thermal printer: 1 unit
- User manual (English + Vietnamese): 1 set
III. Technical Specifications:
- Main machine:
- Examination field: for gynecology, obstetrics, abdominal, small parts, breast, vascular, pediatric, cardiovascular, rectal, neurological, musculoskeletal examinations
- System specifications:
+ Display: LCD/LED or equivalent; Size: ≥ 23 inch; Resolution from 1920 x 1080 pixels or higher
+ Touch screen: ≥ 10.1 inches, LCD or equivalent
+ Number of simultaneous probe ports: ≥ 4 ports
+ Number of digital processing channels: ≥ 1,714,833 channels - maximum image depth: ≥ 42 cm - minimum image depth: ≤ 1 cm
+ Gray scale display: ≥ 256 levels
+ 2D frame rate: ≥ 2800 frames/second
+ Doppler frame rate: ≥ 450 frames/second
+ Volume scanning speed: ≥ 46 frames/second
+ Dynamic range: ≥ 265 dB
+ Operating system: Windows® 10 IoT Enterprise 64-bit or equivalent
- Modes of operation:
- B-mode (2D)
- M-mode
- High repetition frequency Doppler pulse mode
- Color Doppler mode
- Power Doppler mode
- High sensitivity power Doppler mode
- Continuous Doppler mode
- 3D/4D, including static 3D images, real-time 4D images
- Software support for optimizing and image processing:
- Automatic image quality optimization
- Automatic optimization of individual gain
- Automatic PRF adjustment in color Doppler and pulse Doppler
- Cross-beam combination to enhance the clarity of tissue boundaries
- Noise reduction, enhanced smoothness and ultrasound image quality.
- Inverse pulse technology for sharper ultrasound images.
- Real-time ultrasound imaging representation
- Extend observation range for linear and sector probes
- User support software:
- Allows users to reprocess and optimize stored images without requiring the patient; Connects with hospital image storage and management system according to Dicom standard; Automatically calculates real-time Doppler
- Functions include measurement and analysis packages and calculations:
- Basic measurements, PW mode measurements, M-mode measurements, gynecology, fetal ultrasound and Z-scores, pelvic ultrasound, pediatric ultrasound, abdominal ultrasound, small parts ultrasound, breast ultrasound, rectal ultrasound, vascular ultrasound, cardiovascular ultrasound.
- General vascular ultrasound software:
- High sensitivity color Doppler, easily captures low velocity flows
- High sensitivity and resolution Doppler ultrasound technology, blood flow images are as realistic as real blood flow
- 3D/4D Ultrasound Software:
- 3D/4D Ultrasound Software. Automatically adjust the most suitable ROI frame for the best 3D/4D image.
- Technology allows users to adjust the light source position, significantly improving three-dimensional images.
- Cross-sectional scanning - Free drawing of cross sections in 3D/4D data volume '- B Mode Specifications:
- Scanning speed: > 2800 frames/second
- Weak signal filtering: ≥ 51 steps
- Color map: ≥ 10 types
- Gray scale: ≥ 18 types
- M Mode Specifications:
- Scanning speed: ≥ 6 steps
- Storage time: > 60 seconds
- Gray scale: ≥ 18 types
- Color map: ≥ 10 types
- CFM (Color Flow Mapping) Specifications:
- Filter: ≥ 8 steps
- Fine filter: ≥ 12 steps
- Color map: ≥ 8 types
- Repetition frequency: ≤ 100 Hz to ≥ 20.5 kHz
- Color Doppler scanning speed: > 450 frames/second
- PD (Power Doppler) Specifications:
- Filter: ≥ 8 steps
- Fine filter: ≥ 12 steps
- Color map: ≥ 8 levels
- Repetition frequency: ≥ 100 Hz to ≥ 20.5 kHz
- PW (Pulse Wave) Specifications:
- Repetition frequency PRF: PW from ≤ 0.9 kHz to ≥ 22 kHz
- Window width: 0.7, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15 mm
- Color map: ≥ 11 types
- Gray scale: ≥ 18 types
- Continuous Doppler Specifications:
- Maximum flow rate measurement ≥ 30.8 m/s
- 3D/4D, including static 3D images, real-time 4D images:
- Can customize image quality ≥ 6 levels
- Volume scanning speed: ≥ 46 frames/second
- Cine capacity: maximum ≥ 400 frames, ≥ 512 MB
- Minimum connectivity:
- HDMI or equivalent port
- USB connection or compatible port
- Ethernet (RJ45) or compatible connection
- Can be used with biopsy probe
- Accessories:
- Keyboard + mouse: 1 set
- ≥ 04 external covers, continuous ink system or equivalent |
Quynh Luu General Hospital |
. |
. |
|
2 |
Ultrasound |
1 |
Machine |
I. General Requirements:
- The main machine must be manufactured from 2026 onwards, and must be 100% new.
- The origin of the main machine must be from a country in the OECD or G7 group.
- Must meet quality standards: ISO 13485, FDA, or equivalent.
- Power supply must be compatible with Vietnamese power supply: 100 – 240 VAC ~ 50/60 Hz.
- Operating environment:
- Temperature: ≥ 30°C
- Humidity: ≥ 75%
II. Basic Configuration:
Standard configuration for 1 machine:
- 1 main machine designed as a trolley.
- 1 main screen.
- 1 touch control screen.
- 4 probes, including:
- Convex multi-frequency probe.
- 3D/4D volume convex probe.
- Multi-frequency linear probe.
- Multi-frequency microconvex probe.
- Computer system:
- HP desktop computer, processor: ≥ Core i5, minimum RAM: ≥ 8GB, SSD: ≥ 512 GB.
- HP monitor: ≥ 21.5 inches.
- Online UPS ≥ 1KVA.
- 1 thermal printer compatible with the system.
- 4 color printers compatible with the system using water-based ink.
- User manual (English + Vietnamese): 1 set
III. Technical Requirements:
1. Main Machine:
- Must be used for a minimum of the following specialties: obstetrics, gynecology, abdominal, small parts, breast, vascular, pediatrics, cardiology, rectum, neurology, musculoskeletal.
System specifications:
- Display screen: ≥ 21.5 inches LCD + Full HD or higher resolution.
- Touchscreen: ≥ 14 inches LCD + Minimum 03 simultaneous probe ports.
- Digital processing channels: ≥ 9,303,857 channels.
- Scan depth: minimum ≤ 1cm, maximum: ≥ 42 cm.
- Gray scale: ≥ 256 levels.
- 2D frame rate (frames per second): >3000 fps.
- Doppler frame rate: maximum ≥ 740 fps.
- Volume scanning speed: >46 fps.
- Dynamic range: up to ≥ 381 dB.
- Operating system: Windows® 10 IoT Enterprise 64 bit or equivalent.
- Ultrasound modes:
- B-mode (2D)
- M-mode
- 2-cursor simultaneous anatomical M-mode
- High repetition frequency Doppler pulse mode.
- Color Doppler mode.
- Power Doppler mode.
- High sensitivity power Doppler mode.
- 3D/4D mode with static 3D images, real-time 4D images.
Software support for image optimization and processing:
- Automatic image quality optimization with a single button click.
- Automatic gain optimization per section.
- Automatic PRF frequency adjustment in color Doppler and pulse Doppler.
- Real-time B-mode optimization.
- Create combined images from multiple scans taken at different angles into a single image.
- Noise reduction.
- Harmonic coding imaging technique using reverse pulse technology.
- Image B-mode with maximum 22 times magnification.
- Extend the observation field for linear and sector probes.
- Adjust the maximum angle of view on the probe.
Software support for user:
- Automatic real-time Doppler calculation software.
- Allows users to reprocess and optimize stored images without requiring patients.
- Connect with hospital image storage and management system according to Dicom standard.
Software support for gynecology:
- Automatically measure gynecological morphological parameters (BPD, HC, AC, FL, HL) semi-automatically.
- Automatically measure fetal head circumference semi-automatically.
- Automatically measure intracranial diameter semi-automatically.
- Automatically calculate fetal heart rate.
Software support for in-depth fetal cardiac ultrasound:
- M-mode anatomical ultrasound software that allows free angle adjustment of the M, making it easier to obtain optimal images on M-mode.
- AI-based software to help identify fetal cardiac cross-sections.
- General Doppler ultrasound software:
- High sensitivity Doppler technology, blood flow imaging with true blood flow appearance.
- Pre-set settings for optimizing various vascular scanning: uterine artery, middle cerebral artery, umbilical artery, etc.
Software support for 3D/4D ultrasound:
- 3D/4D ultrasound software that automatically adjusts the appropriate ROI frame to create the most beautiful 3D/4D image.
- Technology allows users to adjust the light source position, significantly improving three-dimensional images.
- Fetal cardiac ultrasound - Scan fetal cross-sections freely according to 3D/4D data.
Software support for pelvic ultrasound:
- Pelvic ultrasound software that supports IOTA standard with all 3 criteria LR2, Simple Rule, and ADX.
- Supports scanning of deep endometrial cavity lesions.
- Supports scanning and reporting analysis of uterine wall tumors.
- Measurement and analysis functions: Includes all basic measurements on M-mode, PW mode, gynecology, fetal heart rate, etc.
- Operating modes:
- 3D/4D Mode specifications: Can customize image quality ≥ 6 levels.
- Volume scanning speed: > 46 fps.
- Cine capacity: maximum ≥ 400 frames, ≥ 512 MB.
- B Mode specifications:
- Free scan frame rate according to 3D/4D data
- Scan frame rate: > 3000 fps.
- Weak signal filtering: ≥ 51 steps.
- Color map: ≥ 11 types.
- Gray scale: ≥ 21 types.
- Doppler Mode specifications:
- Pulse repetition frequency (PRF): PW from ≤ 0.9 kHz to ≥ 22 kHz.
- Color Doppler scanning speed: ≥ 740 fps.
- Power Doppler mode specifications:
- Filtering: ≥ 8 steps.
- Smoothing: ≥ 12 steps.
- Color map: ≥ 8 types.
- Pulse repetition frequency (PRF): ≤ 100 Hz to ≥ 20.5 kHz.
- Connectivity:
- HDMI or higher port.
- USB or compatible connection.
- Ethernet (RJ45) or compatible connection.
- Compatible with biopsy probe.
IV. Other Requirements:
- The device must be delivered, installed, and user-trained at the user unit.
- Commitment to provide quality certificate (CQ), origin certificate (CO), customs declaration, and other documents for goods.
- Device warranty period from delivery and acceptance into use: ≥ 24 months according to manufacturer's standards.
- Commitment to supply consumables and accessories within a minimum of 8 years after sales. |
Quynh Luu General Hospital |
. |
. |
|
3 |
DR Image Conversion System |
2 |
System |
I. General Requirements:
- New device, manufactured after 2025
- Origin: OECD member countries
- Meets quality standards: ISO 13485, FDA or equivalent
- Power supply compatible with Vietnamese power grid: ≥ AC 100-240 V~ 50/60 Hz, 1.4-0.7A
- Operating environment:
- Temperature: ≥ 35°C
- Humidity: ≥ 75%
II. Basic Configuration:
- Image acquisition unit: 2 units
- Control box: 1 set
- Batteries: 2 batteries with accessories
- Power cable: 1 set
- Workstation: 1 set:
- Included keyboard and mouse: 1 set
- Image processing software: 1 set
- User manual (English + Vietnamese): 1 set
III. Technical Requirements:
- Wireless image acquisition unit
- Application: General radiography
- TFT type: a-Si or equivalent
- Light source material: CsI
- Dimensions (DxWxH): ≥ 460 mm x 460 mm x 15.6 mm
- Pixel number: ≥ 3072x3072 pixels
- Resolution: ≥ 3.6 lp/mm
- Pixel size: ≥ 140 μm
- TFT operating area (mm): ≥ 430.08 mm × 430.08 mm
- MTF @ 0.5lp/mm: ≥ 84%
- A/D conversion: ≥ 16 bit
- Data output: ≥ 16 bit
- Communication type: Wireless / Wired
- Automatic beam detection: Yes
- Total load capacity: ≥ 400 kg
- Single point load capacity: ≥ 200 kg
- Water and dust protection standard: ≥ IP68
- Sensor protective cover: Carbon fiber
- Weight (kg): ≥ 3.6 kg, including batteries
- Image storage capacity: ≥ 200 images
- Cycle time: maximum ≥ 5 seconds
- Image transfer time: ≥ 1.5 seconds (Wired) and ≥ 2 seconds (Wireless)
- Battery operating time: ≥ 7.5 hours
- Power consumption: ≥ 38W
- Battery capacity: ≥ 4725mAh
- Network connectivity: Supports internet connection according to wireless network standards: ≥ 802.11 a/b/g/n/ac, ≥ 150 Mbps; Wired LAN: ≥ 500 Mbps
- Workstation computer: CPU: ≥ Core i5 (14 cores), RAM: ≥ 8GB, Storage: ≥ 512 GB, OS: Windows 11 or higher
- Included keyboard and mouse
- Screen:
- Size: ≥ 27 inch
- Resolution: ≥ 1920x1080 pixels
- Image processing software with the following tools:
- Patient information search and viewing
- Patient examination record registration / editing / deletion
- Emergency recording mode
- Integration support with DB or MWL
- Image rejection list
- Queue management
- Body part identification through indication
- Supported imaging:
- Minimum software installation: Body map, Exam and Status
- Images obtained through integration of the acquisition unit and the detector
- Preview images
- SW Grid
- Automatic ROI size
- Image viewing functions:
- Post-imaging image processing
- Providing information about the examination
- Zoom / Shrink
- Full screen, fit to screen
- Rotate, Flip, Mirror
- Window/Level
- Text annotation
- Marking
- Inversion
- Graphs
- Measurement and annotation tools:
- Arrows
- Straight lines
- Squares
- Ellipses
- Ruler
- Angle measurement
- Data processing and output functions:
- Image processing
- Select All, Delete
- Image export (Print, CD, USB)
- Auto ROI
- PACS, MWL integration configuration
- DICOM standard image support
IV. Other Requirements:
- The device must be delivered, installed, and user-trained at the user unit.
- Commitment to provide quality certificate (CQ), origin certificate (CO), customs declaration, and other relevant documents for goods.
- Device warranty period from delivery and acceptance into use: ≥ 24 months according to manufacturer's standards.
- Commitment to provide consumables and replacement accessories for a minimum of 8 years after sale. |
Quynh Luu General Hospital |
. |
. |
|
4 |
Electrocardiogram |
2 |
Machine |
I. General Requirements:
- Year of manufacture: 2026 or later, 100% new
- Meets ISO 13485 or equivalent quality standards.
- Power supply compatible with Vietnamese power grid: AC 100-240 V~ 50/60 Hz
- Operating conditions:
+ Ambient temperature: ≥ 30°C
+ Ambient humidity: ≥ 70%
- Origin: Countries within the G7 group
II. Basic Configuration:
- Main unit: 1
- Accessories:
- ECG lead cable: 1
- Power cable: 1
- Chest electrodes: 6
- Limb clip electrodes: 4
- Cart: 1 (to be purchased in Vietnam)
- Printed materials: 1 set
- Cable holder: 1 (to be purchased in Vietnam)
- Battery: 1 set
- User manual (English + Vietnamese): 1 set
III. Technical Specifications:
- General Features
- Data can be output via integrated LAN
- USB or SD card can be used as external storage
- Patient information can be obtained from an external server, and reports can be exported in DICOM or PDF format
- ECG input impedance: ≥ 50 MΩ at 0.67 Hz
- Offset voltage: ± 550 mV
- Input circuit is protected against electrical shock
- Sensitivity: 10 mm/mV, accuracy ±2%
- Common mode rejection ratio (CMRR): > 105 dB
- Noise in: ≤ 20 µVp-p
- Noise between channels: ≤ -40 dB
- Sampling rate: 16,000 samples/second/channel
- Heart rate measurement and display range: 30 – 300 bpm
- Accuracy: ± 10% (30-240 bpm)
- Waveform data processing:
- ECG leads: 12 leads
- Number of input channels: 1 channel
- Built-in AC filter: ≥ 50/60 Hz
- Time constant: ≥ 3.2s
- Sensitivity: 5, 10, 20 mm/mV
- R-wave detection: Yes
- Analog-to-digital converter (ADC): ≥ 24 bits
- Display: TFT LCD or higher
- Size: ≥ 8 inch; Resolution: (800 x 480) points or higher
- Data display: 12 ECG waveforms, patient information, recording mode settings, operating mode, heart rate, QRS synchronization marking, error notifications, electrode displacement and noise
- Printer:
- Method: thermal print
- Printable data: ECG waveform, heart rate, lead name, version, date and time, paper speed, sensitivity, filter settings, patient information, measurement information, event marking.
IV. Other Requirements:
- The device must be delivered, installed, and used proficiently at the user unit.
- Commitment to providing quality certificates (CQ), origin certificates (CO), customs declaration documents, and other relevant documents for the goods.
- Warranty period for the device from delivery and acceptance into use: ≥ 24 months according to manufacturer standards.
- Commitment to provide consumables and replacement parts for a minimum of 8 years after sale. |
Quynh Luu General Hospital |
. |
. |
|
5 |
Gastrointestinal endoscopy system |
1 |
System |
I. General Requirements:
- Manufacturing year: 2026 or later, 100% new
- Meets ISO 13485 or equivalent quality standards.
- Origin: Countries within the G7 group
- Operating conditions:
+ Ambient temperature: ≥ 30°C
+ Ambient humidity: ≥ 70%
- User manual (English + Vietnamese): 1 set
II. Basic Configuration:
- Integrated image processing unit with light source: 1 unit
- Monitor: 1, Full HD or higher resolution
- Video endoscope for stomach and colon, with standard accessories: 1 each
- Video endoscope for colon, with standard accessories: 1 each
- Leak detection device: 1 unit
- System trolley: 1 unit
- Result printing system: 1 set, including:
+ Computer: 1,
+ LCD or higher resolution monitor: 1
+ Compatible printer: 1
- 2-chamber suction device: 1 unit
- Disposable biopsy forceps for stomach endoscope: 1 box
- Disposable biopsy forceps for colon endoscope: 1 box
III. Technical Specifications:
- Integrated image processing unit with light source:
- LED or higher resolution integrated image processing unit
- Provides high-resolution HDTV or higher images
- Supports optical imaging technology in narrow wavelength ranges or blue light mode to improve visualization of blood vessels and other structures
- Output signal compatible with: HDMI, DVI or SDI / HD-SDI / 3G-SDI
- Color adjustment:
- Red: ≥ 9 steps
- Green: ≥ 9 steps
- Chroma: ≥ 9 steps
- Contrast: ≥ 3 levels
- Freeze function
- Adjustable endoscope image size.
- Displays patient information, data storage status, and image information.
- Monitor:
- Medical monitor ≥ 23 inches, Full HD or higher resolution
- Color: ≥ 16 million colors
- Viewing angle: ≥ 89 degrees per side
- Stomach endoscope:
- Supports optical imaging technology in narrow wavelength ranges or short wavelength light generated from optical filters to visualize blood vessels and surfaces with high contrast.
- Field of view ≥ 140°
- Straight viewing direction
- Viewing depth: 2 - ≥ 100 mm
- Outer diameter of the end of the endoscope: ≤ 9.2 mm
- Outer diameter of the body of the endoscope: ≤ 9.2 mm
- Inner diameter of the instrument channel: ≥ 2.8 mm
- Bend angle of the end of the endoscope:
+ Upward: ≥ 210°
+ Downward: ≥ 90°
+ Right: ≥ 100°
+ Left: ≥ 100°
- Working length: ≥ 1000 mm
- Total length: ≥ 1350 mm
- Colon endoscope:
- Supports optical imaging technology in narrow wavelength ranges or short wavelength light generated from optical filters to visualize blood vessels and surfaces with high contrast.
- Field of view: ≥ 140°
- Straight viewing direction
- Viewing depth: 2 - ≥100 mm
- Outer diameter of the end of the endoscope: ≤ 12.8 mm
- Outer diameter of the body of the endoscope: ≤ 12.8 mm
- Inner diameter of the instrument channel: ≥ 3.7 mm
- Bend angle of the end of the endoscope:
+ Upward: ≥ 180°
+ Downward: ≥ 180°
+ Right: ≥ 160°
+ Left: ≥ 160°
- Working length: ≥ 1300 mm
- Total length: ≥ 1650 mm
- Adjustable flexibility of the probe.
- Leak detection device:
- Compatible with the above endoscopes, used for leak testing.
- System trolley:
- Design and structure suitable for the system.
- Includes screen holder, probe holder.
- Four-wheel caster with brake.
- Result printing system:
- Computer with minimum configuration:
+ Core i3 or higher processor; RAM ≥ 4 GB
+ Storage ≥ 350 GB, 17" LCD monitor
+ Mouse, keyboard included
- 2-chamber suction device - Type: oil-free piston pump
- Maximum free air flow rate: ≥ 60 liters/minute
- Noise level: ≤ 55 dB
- Disposable biopsy forceps for stomach endoscope:
- Minimum channel diameter 2.8 mm, oval shape, working length ≥ 1500 mm
- Disposable biopsy forceps for colon endoscope:
- Minimum channel diameter 2.8 mm, oval shape, working length ≥ 2300 mm
IV. Other Requirements:
- The device must be delivered, installed, and user-trained at the user unit.
- Commitment to providing quality certificates (CQ), origin certificates, customs declaration, and other relevant documents.
- Device warranty period from delivery and acceptance for use: ≥ 24 months according to manufacturer's standards.
- Commitment to provide consumables and accessories within a minimum of 8 years after sale. |
Quynh Luu General Hospital |
. |
. |
|
6 |
Endoscopic surgical system |
1 |
System |
I. General Requirements:
- New device, manufactured from 2026 onwards
- Meets ISO 13485 or equivalent quality management standards
- Power supply compatible with Vietnamese power grid: ≥ 100-240 V~ 50/60 Hz
- Operating conditions:
+ Ambient temperature: ≥ 30°C
+ Ambient humidity: ≥ 70%
- Origin: Countries within the G7 group
II. Basic Configuration:
- Main machine system: 1 set, including:
- Image processing unit: 1 piece
- Camera head: 1 piece
- Internal illumination light source (separate or integrated with image processing unit): 1 piece
- Light guide cable: 1 piece
- Dedicated Full HD monitor: 1 set
- High-frequency electrosurgical knife: 1 piece, including:
+ Foot pedal: 1 piece
+ Neutral electrode base: 1 box
+ Connecting cable for neutral electrode base: 1 piece
- CO2 gas pump system: 1 set, including:
+ Gas pumping tube set: 1 set
+ Suction tube set: 1 set
- Operating cart: 1 piece
- User manual (English + Vietnamese): 1 set
III. Technical Specifications:
- Image processing unit:
+ Compatible with both rigid and flexible endoscopes
+ Special observation mode: Narrow band light observation, created from optical filters or infrared light observation
+ ≥ 1 camera connection port
+ Electronic magnification ≥ 1.5 x, with ≥ 3 adjustment levels
+ Image rotation of 180 degrees
+ Color adjustment
+ Contrast: ≥ 3 levels
+ Image enhancement function: including structural and image edge enhancement
+ ≥ 1 USB port
+ Minimum output signal compatible with DVI or equivalent
+ Maximum ≥ 20 user settings and ≥ 50 patient data storage capacity
- Full HD camera head:
+ ≥ 3 CMOS image sensors
+ Resolution: ≥ 1920 x 1080 pixels
+ Electronic or optical magnification: ≥ 2x
+ Camera head focal length range: ≤ 16 - ≥ 30 mm
+ ≥ 2 camera buttons that can be programmed
+ Camera head can be immersed or sterilized
- Internal illumination light source (separate or integrated with image processing unit):
+ Uses LED or equivalent lighting technology
+ Can adjust brightness automatically or manually
+ Brightness adjustment ≥ 17 levels
+ Lamp lifespan: ≥ 10,000 hours
- Light guide cable:
+ Diameter ≥ 4 mm, length ≥ 250 cm
+ Sterilizable
- Dedicated Full HD monitor:
+ LCD or equivalent display
+ Display size: ≥ 27 inches
+ Resolution: ≥ 1920 x 1080 pixels
+ Brightness: ≥ 300 cd/m2
+ Contrast: ≥ 1000:1
+ Viewing angle: ≥ 89 degrees
+ Number of colors displayed: ≥ 16 million colors
+ Minimum signal compatibility with DVI-D or equivalent
- High-frequency electrosurgical knife:
+ Includes all single and bipolar modes for open surgery, endoscopic surgery, and endoscopic intervention
+ Touchscreen for selecting settings
+ ≥ 2 single-pole and ≥ 1 bipolar connection ports
+ Bipolar mode in saltwater environment
+ Electrosurgical knife combined with ultrasonic welding (integrated or separate)
+ Maximum power: ≥ 300W
+ Maximum frequency: ≥ 350 kHz
+ Single-pole function: ≥ 4 single-cut modes, ≥ 4 single-burn modes
+ Bipolar function: ≥ 2 bipolar cut modes, ≥ 5 bipolar burn modes
- CO2 gas pump: Can display the following parameters:
+ Gas pumping mode,
+ Actual pumping pressure,
+ Set pumping pressure,
+ Actual flow rate,
+ Set flow rate
+ Automatic flow rate adjustment function
+ Maximum pumping flow rate ≥ 40 L/min
+ Maximum pumping pressure ≥ 25 mmHg
- Operating cart:
+ Includes 4 wheels with locks
+ ≥ 6 power outlets
+ Movable screen arm |
Quynh Luu General Hospital |
. |
. |
Downloading files directly on the new Public Procurement System requires a computer using Windows operating system and installing Client Agent software (Linux and MacOS operating systems cannot install Client software yet). Therefore, to be able to download files on smartphones, tablets or computers using operating systems other than Windows, you need to use DauThau.info software to download files.
To download, please Login or Register
Downloading files directly on the new Public Procurement System requires a computer using Windows operating system and installing Client Agent software (Linux and MacOS operating systems cannot install Client software yet). Therefore, to be able to download files on smartphones, tablets or computers using operating systems other than Windows, you need to use DauThau.info software to download files.
To download, please Login or Register
- RQ2600032827-01 - Change:Request for quotation code (View changes)
- RQ2600032827-00 Posted for the first time (View changes)
Select columns to display
If you are still not our member, please sign up.You can use your account on DauThau.info to log in to Dauthau.Net, and vice versa!
Only 5 minutes and absolutely free! Your account can be use on all over our ecosystem, including DauThau.info and DauThau.Net.
Detailed instructions on how to deposit money and redeem points on DauThau.info
When does contract negotiation fail?
[VIDEO] Guide for bid solicitors to Extend the Bid Submission Deadline
[VIDEO] Guide to Evaluating Bid Documents on the New National Bidding Network System
[VIDEO] Guide to Posting Tender Notices for PPP Projects
[VIDEO] Guide to Posting Prequalification Notices for PPP Projects
[VIDEO] Guide to Responding to Clarification Requests for e-HSMST (E-Prequalification Invitation Documents) of PPP Projects on the New National Bidding Network System
[VIDEO] Guide to Uploading the Shortlist for PPP Projects
Similar goods purchased by Investor Bệnh viện Đa khoa Quỳnh Lưu:
No similar goods purchased by the Investor were found.
Similar goods purchased by other Investor excluding Bệnh viện Đa khoa Quỳnh Lưu:
No similar goods have been found that other Investor have purchased.
Similar goods for which the bidding results of the Ministry of Health have been announced:
No similar goods have been found in the bidding results of the Ministry of Health.